Asphalt Bent Beam Rheometer (BBR)

قابلة للتفاوضتحديث12/21
نموذج
طبيعة المصنع
المنتجين
فئة المنتج
مكان المنشأ

نظرة عامة

This instrument, the Asphalt Bending Beam Rheometer (BBR), is based on the industry standards JTG E20 "Test Code for Asphalt and Asphalt Mixtures in Highway Engineering" of the People's Republic of China, including T 0627 Asphalt Bending Creep Stiffness Test (Bending Beam Rheometer Method) and SH/T 0775 Asphalt Bending Creep Stiffness Determination Method (BBR Method); Refer to AASHTO T313, AASHTO TP112, and ASTM D6648 of the American Society for Testing and Materials; Designed and manufactured according to European standards such as EN 14771.

تفاصيل المنتج

Asphalt Bent Beam Rheometer (BBR)Suitable for determining the low-temperature bending creep stiffness of asphaltSt)And creep slopemValue can be used to evaluate the low-temperature fracture characteristics of asphalt binders. This instrument is suitable for original asphalt, pressure aged asphalt, and aged asphalt after film oven (or rotary film oven).


1Asphalt Bent Beam Rheometer (BBR)Main technical features and functions

1. The instrument control unit adopts an industrial tablet computer, which displays deformation curves, pressure curves, and temperature curves in real time. The instrument clock software can be set and adjusted.

2. AdoptingPELTIETechnology and refrigeration combined mode, fast cooling speed,40It can be reduced from room temperature to room temperature in minutes-18℃, high temperature control accuracy.

3. Adopting linear differential displacement sensor(LVDT)Equipped with a standard calibration module, displacement parameters can be calibrated accurately and with high precision.

4. Adopting a high-precision, low-noise, and split refrigeration unit, and connecting it to the main unit with a hose to prevent its vibration from affecting the test results.

5. Adopting high-precision load sensors and equipped with standard pressure calibration devices, the pressure can be calibrated accurately and reliably.

6. The host liquid bath is equipped with a magnetic stirrer with adjustable speed, ensuring accurate and uniform temperature of the liquid bath.

7. The air source pressure, pre pressure, load pressure, and air bearing pressure can all be visually displayed in the operating software for easy viewing.

8. Design hasUSBCommunication port, capable of exporting instrument detection data.

9. It has functions such as instrument overload protection and fault warning. When an abnormality occurs, the buzzer will sound an alarm and the interface fault code will be displayed simultaneously.

10. The test time can be set for easy verification240sThe deformation situation after (default)240s).

11. The time required for calculating the results can be set by software (default)8s,15s,30s,60s,120s,240s).

12. Adopting two-stage high-precision air filters to provide high-quality compressed air for the air suspension system, ensuring detection accuracy.


IIMain technical indicators and parameters

1. Host sensor

Displacement sensor range:0~10)mm

Resolution of displacement sensor:5μm

Range of pressure sensor:0~4900)mN

Resolution of pressure sensor:2mN

Temperature sensor range:-50℃~Room temperature

Temperature sensor accuracy:±0.1℃

2. Host liquid bath

Temperature range of liquid bath:-40℃~Room temperature

Temperature control accuracy: ±0.1℃

Liquid bath volume:5L

Stirring speed:25turn/minute

3Other hosts

Trial mold size: 127mm x 12.7mm x 6.35mm

Top radius of sample support: 3mm ± 0.3mm

Loading axis radius: 6.3mm ± 1.3mm spherical contact point

Data interface: USB

Working power supply: AC220V/50Hz; 1kW

Environmental temperature: Room temperature 25 ℃ (air conditioning must be used if necessary)

Environmental humidity: ≤ 70% (relative humidity)

Size: 610mm × 610mm × 620mm

Weight: 80kg